The pneumatic tube system does not affect complete blood count results; a validation study at a tertiary care hospital

A. Z. Al-Riyami, M. Al-Khabori, R. M. Al-Hadhrami, I. S. Al-Azwani, H. M. Davis, K. S. Al-Farsi, S. S. Alkindi, S. F. Daar

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

Introduction: Effect of the pneumatic tube system (PTS) on sample quality is controversial. Herein we aim at evaluating the impact of sample transportation via the PTS on complete blood count (CBC) results. Methods: Duplicate CBC samples from normal donors and anemic patients were sent in parallel to the laboratory for testing through the PTS and the courier (CO). We used scatter plots, Bland-Altman plots, correlation coefficient (r), and coefficient of determination for the validation. Results: A total of 115 samples (donors: 59, patients: 56) were tested. There was excellent correlation between both methods for red blood cell parameters (r range = 0.9213-0.9958) and platelet count. White blood cell (WBC) count and differential count showed similar results (r range = 0.8605-0.9821) for all, with exception of basophils which showed modest correlation (r = 0.4827 for patients and 0.5758 for normal donors). Most of the differences in measurement of all CBC parameters were within the 95% confidence interval of the mean difference on Bland-Altman plots. Conclusion: Modern PTS can be safely used for transporting CBC samples.

Original languageEnglish
Pages (from-to)514-520
Number of pages7
JournalInternational Journal of Laboratory Hematology
Volume36
Issue number5
DOIs
Publication statusPublished - Oct 1 2014

Fingerprint

Pneumatic tubes
Blood Cell Count
Validation Studies
Tertiary Healthcare
Tertiary Care Centers
Blood
Tissue Donors
Basophils
Platelet Count
Leukocyte Count
Cells
Erythrocytes
Confidence Intervals
Platelets
Testing

Keywords

  • Complete blood count
  • Courier
  • Pneumatic tube
  • Quality
  • Transportation

ASJC Scopus subject areas

  • Clinical Biochemistry
  • Biochemistry, medical
  • Hematology
  • Medicine(all)

Cite this

The pneumatic tube system does not affect complete blood count results; a validation study at a tertiary care hospital. / Al-Riyami, A. Z.; Al-Khabori, M.; Al-Hadhrami, R. M.; Al-Azwani, I. S.; Davis, H. M.; Al-Farsi, K. S.; Alkindi, S. S.; Daar, S. F.

In: International Journal of Laboratory Hematology, Vol. 36, No. 5, 01.10.2014, p. 514-520.

Research output: Contribution to journalArticle

@article{4667b2a78d9742bd934129280b627ffe,
title = "The pneumatic tube system does not affect complete blood count results; a validation study at a tertiary care hospital",
abstract = "Introduction: Effect of the pneumatic tube system (PTS) on sample quality is controversial. Herein we aim at evaluating the impact of sample transportation via the PTS on complete blood count (CBC) results. Methods: Duplicate CBC samples from normal donors and anemic patients were sent in parallel to the laboratory for testing through the PTS and the courier (CO). We used scatter plots, Bland-Altman plots, correlation coefficient (r), and coefficient of determination for the validation. Results: A total of 115 samples (donors: 59, patients: 56) were tested. There was excellent correlation between both methods for red blood cell parameters (r range = 0.9213-0.9958) and platelet count. White blood cell (WBC) count and differential count showed similar results (r range = 0.8605-0.9821) for all, with exception of basophils which showed modest correlation (r = 0.4827 for patients and 0.5758 for normal donors). Most of the differences in measurement of all CBC parameters were within the 95{\%} confidence interval of the mean difference on Bland-Altman plots. Conclusion: Modern PTS can be safely used for transporting CBC samples.",
keywords = "Complete blood count, Courier, Pneumatic tube, Quality, Transportation",
author = "Al-Riyami, {A. Z.} and M. Al-Khabori and Al-Hadhrami, {R. M.} and Al-Azwani, {I. S.} and Davis, {H. M.} and Al-Farsi, {K. S.} and Alkindi, {S. S.} and Daar, {S. F.}",
year = "2014",
month = "10",
day = "1",
doi = "10.1111/ijlh.12180",
language = "English",
volume = "36",
pages = "514--520",
journal = "International Journal of Laboratory Hematology",
issn = "1751-5521",
publisher = "Wiley-Blackwell",
number = "5",

}

TY - JOUR

T1 - The pneumatic tube system does not affect complete blood count results; a validation study at a tertiary care hospital

AU - Al-Riyami, A. Z.

AU - Al-Khabori, M.

AU - Al-Hadhrami, R. M.

AU - Al-Azwani, I. S.

AU - Davis, H. M.

AU - Al-Farsi, K. S.

AU - Alkindi, S. S.

AU - Daar, S. F.

PY - 2014/10/1

Y1 - 2014/10/1

N2 - Introduction: Effect of the pneumatic tube system (PTS) on sample quality is controversial. Herein we aim at evaluating the impact of sample transportation via the PTS on complete blood count (CBC) results. Methods: Duplicate CBC samples from normal donors and anemic patients were sent in parallel to the laboratory for testing through the PTS and the courier (CO). We used scatter plots, Bland-Altman plots, correlation coefficient (r), and coefficient of determination for the validation. Results: A total of 115 samples (donors: 59, patients: 56) were tested. There was excellent correlation between both methods for red blood cell parameters (r range = 0.9213-0.9958) and platelet count. White blood cell (WBC) count and differential count showed similar results (r range = 0.8605-0.9821) for all, with exception of basophils which showed modest correlation (r = 0.4827 for patients and 0.5758 for normal donors). Most of the differences in measurement of all CBC parameters were within the 95% confidence interval of the mean difference on Bland-Altman plots. Conclusion: Modern PTS can be safely used for transporting CBC samples.

AB - Introduction: Effect of the pneumatic tube system (PTS) on sample quality is controversial. Herein we aim at evaluating the impact of sample transportation via the PTS on complete blood count (CBC) results. Methods: Duplicate CBC samples from normal donors and anemic patients were sent in parallel to the laboratory for testing through the PTS and the courier (CO). We used scatter plots, Bland-Altman plots, correlation coefficient (r), and coefficient of determination for the validation. Results: A total of 115 samples (donors: 59, patients: 56) were tested. There was excellent correlation between both methods for red blood cell parameters (r range = 0.9213-0.9958) and platelet count. White blood cell (WBC) count and differential count showed similar results (r range = 0.8605-0.9821) for all, with exception of basophils which showed modest correlation (r = 0.4827 for patients and 0.5758 for normal donors). Most of the differences in measurement of all CBC parameters were within the 95% confidence interval of the mean difference on Bland-Altman plots. Conclusion: Modern PTS can be safely used for transporting CBC samples.

KW - Complete blood count

KW - Courier

KW - Pneumatic tube

KW - Quality

KW - Transportation

UR - http://www.scopus.com/inward/record.url?scp=84908226479&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84908226479&partnerID=8YFLogxK

U2 - 10.1111/ijlh.12180

DO - 10.1111/ijlh.12180

M3 - Article

C2 - 24373139

AN - SCOPUS:84908226479

VL - 36

SP - 514

EP - 520

JO - International Journal of Laboratory Hematology

JF - International Journal of Laboratory Hematology

SN - 1751-5521

IS - 5

ER -