Survey of computed tomography technique and radiation dose in Sudanese hospitals

I. I. Suliman, S. E. Abdalla, Nada A. Ahmed, M. A. Galal, Isam Salih

Research output: Contribution to journalArticle

15 Citations (Scopus)

Abstract

The purpose of this study was to survey technique and radiation absorbed dose in CT examinations of adult in Sudan and to compare the results with the reference dose levels. Questionnaire forms were completed in nine hospitals and a sample of 445 CT examinations in patients. Information on patient, procedure, scanner, and technique for common CT examinations were collected. For each facility, the radiation absorbed dose was measured on CT dose phantom measuring 16 cm (head) and 32 cm (body) in diameter and was used to calculate the normalized CT air kerma index. Volume CT air kerma index (C VOL), CT air kerma-length product (P KL,CT) values were calculated using the measured normalized CT air kerma index and questionnaire information. The effective dose, E estimates was determined by using P KL,CT measurements and appropriate normalized coefficients. Assuming the sample to offer a fair representative picture of CT practice patterns in Sudan, the mean C VOL and P KL,CT values were comparable or below the reference doses: 65 mGy and 758 mGy cm, respectively at head CT; 11.5 mGy and 327 mGy cm, respectively at chest CT; 11.6 mGy and 437 mGy cm, respectively at abdominal CT; and 11.0 mGy and 264 mGy cm, respectively at pelvis CT. Estimated effective doses were 1.6, 4.6, 6.6 and 4.0 mSv, respectively. The study offered a first national dose survey and provided a mean for quality control and optimization of CT practice within the country.

Original languageEnglish
JournalEuropean Journal of Radiology
Volume80
Issue number3
DOIs
Publication statusPublished - Dec 2011

Fingerprint

Air
Tomography
Radiation
Sudan
Head
Cone-Beam Computed Tomography
Pelvis
Quality Control
Thorax
Surveys and Questionnaires
kloxil

Keywords

  • CT
  • CT air kerma dose index
  • CT air kerma-length product
  • Radiation dose

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Survey of computed tomography technique and radiation dose in Sudanese hospitals. / Suliman, I. I.; Abdalla, S. E.; Ahmed, Nada A.; Galal, M. A.; Salih, Isam.

In: European Journal of Radiology, Vol. 80, No. 3, 12.2011.

Research output: Contribution to journalArticle

Suliman, I. I. ; Abdalla, S. E. ; Ahmed, Nada A. ; Galal, M. A. ; Salih, Isam. / Survey of computed tomography technique and radiation dose in Sudanese hospitals. In: European Journal of Radiology. 2011 ; Vol. 80, No. 3.
@article{e1485630d48e4dc38d8b840d9a59e636,
title = "Survey of computed tomography technique and radiation dose in Sudanese hospitals",
abstract = "The purpose of this study was to survey technique and radiation absorbed dose in CT examinations of adult in Sudan and to compare the results with the reference dose levels. Questionnaire forms were completed in nine hospitals and a sample of 445 CT examinations in patients. Information on patient, procedure, scanner, and technique for common CT examinations were collected. For each facility, the radiation absorbed dose was measured on CT dose phantom measuring 16 cm (head) and 32 cm (body) in diameter and was used to calculate the normalized CT air kerma index. Volume CT air kerma index (C VOL), CT air kerma-length product (P KL,CT) values were calculated using the measured normalized CT air kerma index and questionnaire information. The effective dose, E estimates was determined by using P KL,CT measurements and appropriate normalized coefficients. Assuming the sample to offer a fair representative picture of CT practice patterns in Sudan, the mean C VOL and P KL,CT values were comparable or below the reference doses: 65 mGy and 758 mGy cm, respectively at head CT; 11.5 mGy and 327 mGy cm, respectively at chest CT; 11.6 mGy and 437 mGy cm, respectively at abdominal CT; and 11.0 mGy and 264 mGy cm, respectively at pelvis CT. Estimated effective doses were 1.6, 4.6, 6.6 and 4.0 mSv, respectively. The study offered a first national dose survey and provided a mean for quality control and optimization of CT practice within the country.",
keywords = "CT, CT air kerma dose index, CT air kerma-length product, Radiation dose",
author = "Suliman, {I. I.} and Abdalla, {S. E.} and Ahmed, {Nada A.} and Galal, {M. A.} and Isam Salih",
year = "2011",
month = "12",
doi = "10.1016/j.ejrad.2010.12.050",
language = "English",
volume = "80",
journal = "European Journal of Radiology",
issn = "0720-048X",
publisher = "Elsevier Ireland Ltd",
number = "3",

}

TY - JOUR

T1 - Survey of computed tomography technique and radiation dose in Sudanese hospitals

AU - Suliman, I. I.

AU - Abdalla, S. E.

AU - Ahmed, Nada A.

AU - Galal, M. A.

AU - Salih, Isam

PY - 2011/12

Y1 - 2011/12

N2 - The purpose of this study was to survey technique and radiation absorbed dose in CT examinations of adult in Sudan and to compare the results with the reference dose levels. Questionnaire forms were completed in nine hospitals and a sample of 445 CT examinations in patients. Information on patient, procedure, scanner, and technique for common CT examinations were collected. For each facility, the radiation absorbed dose was measured on CT dose phantom measuring 16 cm (head) and 32 cm (body) in diameter and was used to calculate the normalized CT air kerma index. Volume CT air kerma index (C VOL), CT air kerma-length product (P KL,CT) values were calculated using the measured normalized CT air kerma index and questionnaire information. The effective dose, E estimates was determined by using P KL,CT measurements and appropriate normalized coefficients. Assuming the sample to offer a fair representative picture of CT practice patterns in Sudan, the mean C VOL and P KL,CT values were comparable or below the reference doses: 65 mGy and 758 mGy cm, respectively at head CT; 11.5 mGy and 327 mGy cm, respectively at chest CT; 11.6 mGy and 437 mGy cm, respectively at abdominal CT; and 11.0 mGy and 264 mGy cm, respectively at pelvis CT. Estimated effective doses were 1.6, 4.6, 6.6 and 4.0 mSv, respectively. The study offered a first national dose survey and provided a mean for quality control and optimization of CT practice within the country.

AB - The purpose of this study was to survey technique and radiation absorbed dose in CT examinations of adult in Sudan and to compare the results with the reference dose levels. Questionnaire forms were completed in nine hospitals and a sample of 445 CT examinations in patients. Information on patient, procedure, scanner, and technique for common CT examinations were collected. For each facility, the radiation absorbed dose was measured on CT dose phantom measuring 16 cm (head) and 32 cm (body) in diameter and was used to calculate the normalized CT air kerma index. Volume CT air kerma index (C VOL), CT air kerma-length product (P KL,CT) values were calculated using the measured normalized CT air kerma index and questionnaire information. The effective dose, E estimates was determined by using P KL,CT measurements and appropriate normalized coefficients. Assuming the sample to offer a fair representative picture of CT practice patterns in Sudan, the mean C VOL and P KL,CT values were comparable or below the reference doses: 65 mGy and 758 mGy cm, respectively at head CT; 11.5 mGy and 327 mGy cm, respectively at chest CT; 11.6 mGy and 437 mGy cm, respectively at abdominal CT; and 11.0 mGy and 264 mGy cm, respectively at pelvis CT. Estimated effective doses were 1.6, 4.6, 6.6 and 4.0 mSv, respectively. The study offered a first national dose survey and provided a mean for quality control and optimization of CT practice within the country.

KW - CT

KW - CT air kerma dose index

KW - CT air kerma-length product

KW - Radiation dose

UR - http://www.scopus.com/inward/record.url?scp=80755172161&partnerID=8YFLogxK