Performance testing of selected types of electronic personal dosimeters used in Sudan

I. I. Suliman, E. H. Yousif, A. A. Beineen, B. E. Yousif, M. Hassan

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

Measurements were carried out for calibration and performance testing of a set of 10 electronic personal dosimeters (EPDs) at the Secondary Standard Dosimetry Laboratory of Sudan. Calibrations were carried out at three X-ray beam qualities described in ISO standard 4037 in addition to 137Cs and 60Co gamma ray beams. The experimental was performed with EPDs mounted on ICRU Slab phantom. X-ray and γ-ray beams were characterized in terms of air kerma free-in-air which were converted to the known delivered personal dose equivalent, Hp(10) using appropriate the air kerma to personal dose equivalent conversion coefficients. Dosimeters tested showed excellent energy and angular response and relative error of indication within the recommended limit for photon energies from 65 keV to 1.25 MeV. The study showed encouraging results for using electronic dosimeters in personal dosimetry.

Original languageEnglish
Pages (from-to)1582-1584
Number of pages3
JournalRadiation Measurements
Volume45
Issue number10
DOIs
Publication statusPublished - Dec 2010

Fingerprint

Sudan
Dosimeters
dosimeters
Dosimetry
Testing
electronics
Air
Calibration
X rays
Beam quality
air
Gamma rays
gamma ray beams
dosage
Photons
rays
indication
slabs
x rays
energy

Keywords

  • Electronic personal dosimeters
  • Personal dose equivalent
  • Radiation dosimetry

ASJC Scopus subject areas

  • Radiation
  • Instrumentation

Cite this

Suliman, I. I., Yousif, E. H., Beineen, A. A., Yousif, B. E., & Hassan, M. (2010). Performance testing of selected types of electronic personal dosimeters used in Sudan. Radiation Measurements, 45(10), 1582-1584. https://doi.org/10.1016/j.radmeas.2010.05.032

Performance testing of selected types of electronic personal dosimeters used in Sudan. / Suliman, I. I.; Yousif, E. H.; Beineen, A. A.; Yousif, B. E.; Hassan, M.

In: Radiation Measurements, Vol. 45, No. 10, 12.2010, p. 1582-1584.

Research output: Contribution to journalArticle

Suliman, II, Yousif, EH, Beineen, AA, Yousif, BE & Hassan, M 2010, 'Performance testing of selected types of electronic personal dosimeters used in Sudan', Radiation Measurements, vol. 45, no. 10, pp. 1582-1584. https://doi.org/10.1016/j.radmeas.2010.05.032
Suliman, I. I. ; Yousif, E. H. ; Beineen, A. A. ; Yousif, B. E. ; Hassan, M. / Performance testing of selected types of electronic personal dosimeters used in Sudan. In: Radiation Measurements. 2010 ; Vol. 45, No. 10. pp. 1582-1584.
@article{e041fc6e402648edb6a6db33cc8da593,
title = "Performance testing of selected types of electronic personal dosimeters used in Sudan",
abstract = "Measurements were carried out for calibration and performance testing of a set of 10 electronic personal dosimeters (EPDs) at the Secondary Standard Dosimetry Laboratory of Sudan. Calibrations were carried out at three X-ray beam qualities described in ISO standard 4037 in addition to 137Cs and 60Co gamma ray beams. The experimental was performed with EPDs mounted on ICRU Slab phantom. X-ray and γ-ray beams were characterized in terms of air kerma free-in-air which were converted to the known delivered personal dose equivalent, Hp(10) using appropriate the air kerma to personal dose equivalent conversion coefficients. Dosimeters tested showed excellent energy and angular response and relative error of indication within the recommended limit for photon energies from 65 keV to 1.25 MeV. The study showed encouraging results for using electronic dosimeters in personal dosimetry.",
keywords = "Electronic personal dosimeters, Personal dose equivalent, Radiation dosimetry",
author = "Suliman, {I. I.} and Yousif, {E. H.} and Beineen, {A. A.} and Yousif, {B. E.} and M. Hassan",
year = "2010",
month = "12",
doi = "10.1016/j.radmeas.2010.05.032",
language = "English",
volume = "45",
pages = "1582--1584",
journal = "Radiation Measurements",
issn = "1350-4487",
publisher = "Elsevier Limited",
number = "10",

}

TY - JOUR

T1 - Performance testing of selected types of electronic personal dosimeters used in Sudan

AU - Suliman, I. I.

AU - Yousif, E. H.

AU - Beineen, A. A.

AU - Yousif, B. E.

AU - Hassan, M.

PY - 2010/12

Y1 - 2010/12

N2 - Measurements were carried out for calibration and performance testing of a set of 10 electronic personal dosimeters (EPDs) at the Secondary Standard Dosimetry Laboratory of Sudan. Calibrations were carried out at three X-ray beam qualities described in ISO standard 4037 in addition to 137Cs and 60Co gamma ray beams. The experimental was performed with EPDs mounted on ICRU Slab phantom. X-ray and γ-ray beams were characterized in terms of air kerma free-in-air which were converted to the known delivered personal dose equivalent, Hp(10) using appropriate the air kerma to personal dose equivalent conversion coefficients. Dosimeters tested showed excellent energy and angular response and relative error of indication within the recommended limit for photon energies from 65 keV to 1.25 MeV. The study showed encouraging results for using electronic dosimeters in personal dosimetry.

AB - Measurements were carried out for calibration and performance testing of a set of 10 electronic personal dosimeters (EPDs) at the Secondary Standard Dosimetry Laboratory of Sudan. Calibrations were carried out at three X-ray beam qualities described in ISO standard 4037 in addition to 137Cs and 60Co gamma ray beams. The experimental was performed with EPDs mounted on ICRU Slab phantom. X-ray and γ-ray beams were characterized in terms of air kerma free-in-air which were converted to the known delivered personal dose equivalent, Hp(10) using appropriate the air kerma to personal dose equivalent conversion coefficients. Dosimeters tested showed excellent energy and angular response and relative error of indication within the recommended limit for photon energies from 65 keV to 1.25 MeV. The study showed encouraging results for using electronic dosimeters in personal dosimetry.

KW - Electronic personal dosimeters

KW - Personal dose equivalent

KW - Radiation dosimetry

UR - http://www.scopus.com/inward/record.url?scp=78649784770&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=78649784770&partnerID=8YFLogxK

U2 - 10.1016/j.radmeas.2010.05.032

DO - 10.1016/j.radmeas.2010.05.032

M3 - Article

VL - 45

SP - 1582

EP - 1584

JO - Radiation Measurements

JF - Radiation Measurements

SN - 1350-4487

IS - 10