Effective dose calculations in conventional diagnostic x-ray examinations for patients in major sudanese hospitals

Ibrahim Idris Suliman*, F. I. Habbani

*Corresponding author for this work

Research output: Chapter in Book/Report/Conference proceedingConference contribution

2 Citations (Scopus)

Abstract

The purpose of this work was to compare two methods used for effective dose (ED) calculation in conventional diagnostic x-ray examinations and to analyze effective dose distributions among radiological departments under study. The study was performed in four major hospitals in Khartoum area, Sudan, covering eight x-ray units and a sample of 325 patients. Two approaches were used for effective dose calculation: In the first approach, the entrance surface dose (ESD) values were estimated from the x-ray tube output parameters for chest PA, Skull AP/PA, Skull LAT, Pelvis AP, Lumbar Spine AP and Lumbar Spine LAT examinations. The ED values were then calculated from the obtained ESD values using NRPBSR262 Monte Carlo data and XDOSE software. In the second approach, the energy imparted to patients in the same x-ray examinations was computed using values for entrance skin exposure, half- value layer and exposure area. Effective doses were then calculated from energy imparted using ED conversion factors proposed by other authors. The results of ED values calculated using the two approaches were in good agreement between themselves and with data reported in the literature. The mean ED values calculated using the Monte Carlo data and XDOSE software were found to be 29.9, 14.1, 8.91, 232.5, 224.5 and 93.6 μSv for chest PA, Skull AP/PA, Skull LAT, Pelvis AP, Lumbar Spine AP and Lumbar Spine LAT examinations, respectively. Almost all ED values were below the corresponding DRLs established elsewhere, except for chest PA examination which exhibited comparable mean ED values. The study is expected to increase the awareness among the medical professionals and members of the public on the status of radiological protection in Sudan.

Original languageEnglish
Title of host publicationIFMBE Proceedings
EditorsSun I. Kim, Tae Suk Suh
PublisherSpringer Verlag
Pages2087-2091
Number of pages5
Edition1
ISBN (Print)9783540368397
DOIs
Publication statusPublished - 2007
Externally publishedYes
Event10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 - Seoul, Korea, Republic of
Duration: Aug 27 2006Sept 1 2006

Publication series

NameIFMBE Proceedings
Number1
Volume14
ISSN (Print)1680-0737
ISSN (Electronic)1433-9277

Other

Other10th World Congress on Medical Physics and Biomedical Engineering, WC 2006
Country/TerritoryKorea, Republic of
CitySeoul
Period8/27/069/1/06

Keywords

  • Diagnostic radiology
  • Effective dose
  • Energy imparted

ASJC Scopus subject areas

  • Bioengineering
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Effective dose calculations in conventional diagnostic x-ray examinations for patients in major sudanese hospitals'. Together they form a unique fingerprint.

Cite this