Abstract
A study was performed to compare adult patient doses in film screen (FS) and computed radiography (CR) diagnostic X-ray examinations in some hospitals in Sudan over a period of 1 y; during this period of time, the CR systems were introduced to replace FS systems. Radiation doses were estimated for 354 patients in five hospitals (two FS units and three CR units). Entrance surface air kerma (ESAK) was estimated from incident air kerma using patient exposure parameters and tube output. Dose calculations were performed using CALDOSE X 3.5 Monte Carlo-based software. In FS, third quartile of ESAK values for skull PA, skull LAT, chest PA, pelvis AP, lumbar spine AP and lumbar spine LAT were 1.5, 1.3, 0.3, 1.9, 2.8 and 5.9 mGy, respectively, while in CR, third quartile of ESAKvalues for the same examinations were 2.7, 1.7, 0.18, 1.7, 3.2 and 10.8 mGy, respectively. Comparable ESAK values were presented in FS and CR units. The results are important for future dose optimisation and setting national diagnostic reference levels.
Original language | English |
---|---|
Pages (from-to) | 402-405 |
Number of pages | 4 |
Journal | Radiation Protection Dosimetry |
Volume | 165 |
Issue number | 1-4 |
DOIs | |
Publication status | Published - Jul 1 2015 |
ASJC Scopus subject areas
- Radiological and Ultrasound Technology
- Radiation
- Radiology Nuclear Medicine and imaging
- Public Health, Environmental and Occupational Health