Traumatic Dehiscence of Medtronic Freestyle Stentless Bioprosthesis

Turki Albacker, Adil Al Kindi, Benoit de Varennes*

*المؤلف المقابل لهذا العمل

نتاج البحث: المساهمة في مجلةArticleمراجعة النظراء

4 اقتباسات (Scopus)

ملخص

We report a case of deceleration injury causing traumatic dehiscence of a Medtronic Freestyle aortic bioprosthesis (Medtronic Inc, Minneapolis, MN) in a 70-year-old patient 5 years after a total root reconstruction. The patient presented with progressive exertional dyspnea 6 months after a significant impact during a hockey game. Intraoperatively, there was a posterior dehiscence of the proximal suture line and almost complete tear of the left coronary leaflet. At reoperation, the patient underwent reconstruction of the aortic root and Bentall procedure with a size 23 mechanical aortic valved conduit.

اللغة الأصليةEnglish
الصفحات (من إلى)643-645
عدد الصفحات3
دوريةAnnals of Thoracic Surgery
مستوى الصوت86
رقم الإصدار2
المعرِّفات الرقمية للأشياء
حالة النشرPublished - أغسطس 2008
منشور خارجيًانعم

ASJC Scopus subject areas

  • ???subjectarea.asjc.2700.2746???
  • ???subjectarea.asjc.2700.2740???
  • ???subjectarea.asjc.2700.2705???

بصمة

أدرس بدقة موضوعات البحث “Traumatic Dehiscence of Medtronic Freestyle Stentless Bioprosthesis'. فهما يشكلان معًا بصمة فريدة.

قم بذكر هذا