Photoneutrons and Gamma Capture Dose Rates at the Maze Entrance of Varian TrueBeam and Elekta Versa HD Medical Linear Accelerators

Ibrahim I. Suliman*, Ghada A. Khouqeer, Fareed H. Mayhoub

*المؤلف المقابل لهذا العمل

نتاج البحث: المساهمة في مجلةArticleمراجعة النظراء

1 اقتباس (Scopus)

ملخص

Herein, we evaluated the neutron and gamma capture dose equivalent rates at the maze entrance of Varian TrueBeam and Elekta Versa HD™ medical linear accelerators (linacs) using experimental measurements as well as empirical calculations. Dose rates were measured using calibrated neutron and gamma area survey meters placed side-by-side at the measurement point of interest. Measurements were performed at a source-to-detector distance of 100 cm, with a 10 × 10 cm2 field size therapeutic X-ray beam, and a 30 × 30 × 15 cm3 solid water patient equivalent phantom, with a linac operating at 15, 10 MV, and 10 MV flattened filter-free (FFF). Dose rates were also measured at different points at the centerline along the maze towards the maze entrance. The measured dose equivalent rates at the maze entrance were comparable to those reported in the literature. The dose rates along the maze decreased exponentially towards the maze entrance and were significant for short maze lengths. The evaluated empirical methods for estimating neutron dose rates at the maze entrance of a linac proposed by Kersey, the modified Kersey method and Falcão method, agree by a factor of two from the experimental measurements. The results revealed vital radiation protection considerations owing to neutron contamination in external beam therapy.

اللغة الأصليةEnglish
رقم المقال78
دوريةToxics
مستوى الصوت11
رقم الإصدار1
المعرِّفات الرقمية للأشياء
حالة النشرPublished - يناير 14 2023

ASJC Scopus subject areas

  • ???subjectarea.asjc.3000.3005???
  • ???subjectarea.asjc.2300.2307???
  • ???subjectarea.asjc.1500.1504???

قم بذكر هذا