Performance testing of selected types of electronic personal dosimeters used in Sudan

I. I. Suliman, E. H. Yousif, A. A. Beineen, B. E. Yousif, M. Hassan

نتاج البحث: المساهمة في مجلةArticleمراجعة النظراء

2 اقتباسات (Scopus)

ملخص

Measurements were carried out for calibration and performance testing of a set of 10 electronic personal dosimeters (EPDs) at the Secondary Standard Dosimetry Laboratory of Sudan. Calibrations were carried out at three X-ray beam qualities described in ISO standard 4037 in addition to 137Cs and 60Co gamma ray beams. The experimental was performed with EPDs mounted on ICRU Slab phantom. X-ray and γ-ray beams were characterized in terms of air kerma free-in-air which were converted to the known delivered personal dose equivalent, Hp(10) using appropriate the air kerma to personal dose equivalent conversion coefficients. Dosimeters tested showed excellent energy and angular response and relative error of indication within the recommended limit for photon energies from 65 keV to 1.25 MeV. The study showed encouraging results for using electronic dosimeters in personal dosimetry.

اللغة الأصليةEnglish
الصفحات (من إلى)1582-1584
عدد الصفحات3
دوريةRadiation Measurements
مستوى الصوت45
رقم الإصدار10
المعرِّفات الرقمية للأشياء
حالة النشرPublished - ديسمبر 2010
منشور خارجيًانعم