Optimisation of T2 and T2* sequences in MRI for better quantification of iron on transfused dependent sickle cell patients

Azza Ahmed*, Amani Baldo, A. Sulieman, Hind Mirghani, Fouad A. Abolaban, I. I. Suliman, Isam Salih*

*المؤلف المقابل لهذا العمل

نتاج البحث: المساهمة في مجلةArticleمراجعة النظراء

1 اقتباس (Scopus)

ملخص

This work aimed to investigate the effect of different shim techniques, voxel sizes, and repetition time (TR) on using theT2 and T2* sequences to determine their optimum settings to investigate the quantification of iron in transfused dependent sickle cell patients. The effect of each of these parameters was investigated on phantoms of different Gadolinium (Gd) concentrations, on 10 volunteers and 25 patients using a1 5T MRI Philips scanner. No significant difference between the three shim techniques was noticed in either T2 or T2* sequence measurements. Pixel sizes of 1 × 1 and 2 × 2 mm provided optimum results for T2 measurements. At 1 × 1 mm pixel size the T2* measurements experienced less error in measurements than the size of 2.5 × 2.5 mm used in the literature. Even though the slice thickness variation did not provide any changes in T2 measurements, the 12 mm provided optimum T2* measurements. TR variation did not yield significant changes on either T2 or T2* measurements. These results indicate that both T2 and T2* sequences can be further improved by providing more reliable measurements and reducing acquisition time.

اللغة الأصليةEnglish
رقم المقال8513
دوريةScientific Reports
مستوى الصوت11
رقم الإصدار1
المعرِّفات الرقمية للأشياء
حالة النشرPublished - ديسمبر 2021
منشور خارجيًانعم

ASJC Scopus subject areas

  • ???subjectarea.asjc.1000???

بصمة

أدرس بدقة موضوعات البحث “Optimisation of T2 and T2* sequences in MRI for better quantification of iron on transfused dependent sickle cell patients'. فهما يشكلان معًا بصمة فريدة.

قم بذكر هذا