Highly Sensitive and Selective Voltammetric Sensor Fullerene Modified Glassy Carbon Electrode for Determination of Cefitizoxime in Solubilized System

Rajeev Jain*, Jahangir Ahmad Rather, Ashish Dwivedi, Vikas

*المؤلف المقابل لهذا العمل

نتاج البحث: المساهمة في مجلةArticleمراجعة النظراء

40 اقتباسات (Scopus)

ملخص

The usefulness of fullerene modified glassy carbon electrode in mediating the reduction of cefitizoxime in solubilized system has been demonstrated. Due to the unique structure and extraordinary properties, fullerene shows higher catalytic efficiency towards cefitizoxime reduction. The kinetic parameters, electron transfer coefficient (α) and rate constant (K0) across the modified electrode are 0.37 and 0.1081/s respectively. The proposed square-wave voltammetric method is linear over the concentration range 1.2-10.3 μg/mL. The limit of detection (LOD) is found 0.27 ng/mL. High sensitivity and selectivity together with low detection limit of the electrode response make it suitable for the determination of cefitizoxime.

اللغة الأصليةEnglish
الصفحات (من إلى)2600-2606
عدد الصفحات7
دوريةElectroanalysis
مستوى الصوت22
رقم الإصدار21
المعرِّفات الرقمية للأشياء
حالة النشرPublished - نوفمبر 2010
منشور خارجيًانعم

ASJC Scopus subject areas

  • ???subjectarea.asjc.1600.1602???
  • ???subjectarea.asjc.1600.1603???

بصمة

أدرس بدقة موضوعات البحث “Highly Sensitive and Selective Voltammetric Sensor Fullerene Modified Glassy Carbon Electrode for Determination of Cefitizoxime in Solubilized System'. فهما يشكلان معًا بصمة فريدة.

قم بذكر هذا